What it does
A conventional generator delivers a carrier with simple modulations. A vector signal generator creates signals like those of real communication systems: filtered digital modulations, frames with preambles, bursts and multiple carriers. To do this it works with a complex signal described by two components, I and Q, which set the amplitude and phase at each instant.
How it is built
- Waveform generation: a processor calculates, or plays back from memory, the I and Q samples of the desired format.
- Digital-to-analog converters: they turn the samples into analog baseband signals.
- I/Q modulator: it mixes both components with a carrier and its phase-shifted version and adds them, which gives the modulated signal at the output frequency.
- Output stage: amplification and controlled attenuation to set the level accurately.
What it is used for
- Receiver testing: sensitivity, selectivity, blocking and error rate with realistic signals.
- Stimulating amplifiers and mixers to measure linearity and leakage into adjacent channels.
- Generating controlled interferers to study coexistence.
- Replaying a signal captured with an analyzer, to repeat a problem observed in the field.
Deliberate impairments
One advantage of the vector approach is that measured defects can be added: noise, I/Q imbalance, frequency or phase errors. This makes it possible to verify how much the receiver can withstand, or to calibrate a measurement system. Conversely, the generator has its own residual defects, and to test a very demanding receiver its signal quality must be clearly better than the quality to be verified.
What to watch for
- The actual output level depends on the calibration and on the peak-to-average ratio of the waveform; if the waveform has high peaks, the generator can compress and distort.
- The loss of cables and adapters must be compensated to know what level reaches the equipment.
- The phase noise of the internal oscillator can limit tests of systems that are sensitive to it.