24/7 emergencies 814 777 6921812 230 6079WhatsApp 811 786 0768
Electrónica MonterreyElectrónica MonterreyLaboratorios Creativeit Talk to an engineer
Menu

How is transmitter output power measured?

The output power of a transmitter is measured by connecting it, through a suitable attenuator, to a power meter or an analyzer, which integrates the power within the signal bandwidth, and correcting for the path losses.

What needs to be known

There are several possible powers, and it is worth stating which one is measured: the average power during transmission, the peak power or the average power during a burst. In signals with complex modulation, the difference between peak and average can be large, and confusing them leads to wrong conclusions about compliance or linearity.

Basic setup

  1. The transmitter output is connected to the instrument through a good quality cable and a power attenuator that protects the input and reduces reflection.
  2. The losses of the attenuator and cables are measured separately at the working frequency and compensated for.
  3. The instrument is set for the frequency and signal type, and the measurement is taken.

Instruments

  • Power meter with sensor: it measures the power of the entire signal entering the sensor, whether it is the desired one or not. It is very accurate for wideband signals, but it does not distinguish what is inside the band.
  • Spectrum or signal analyzer: it allows the power integrated over a defined bandwidth to be measured and harmonics and spurs to be seen; it usually has lower absolute accuracy than a dedicated meter.
  • Meters with peak and burst detection: needed for pulsed or burst signals, so that intervals without transmission are not averaged in.

Common sources of error in the method

  • Impedance mismatch between transmitter and sensor, which changes the power delivered.
  • Dirty or poorly tightened connectors, which alter the loss.
  • Saturation of the instrument input due to lack of attenuation.
  • Failing to consider the duty cycle in pulsed transmissions.
  • Expired calibration of the sensor or analyzer.

How to interpret the result

The value is expressed in decibels relative to one milliwatt or in watts, always referred to the output connector if measured by conduction. If the equipment has an integrated antenna, the radiated power is measured in a controlled environment and related to the antenna gain, which is different from conducted power.

High-power transmitters produce dangerous voltages and radiation. Their measurement must be carried out by qualified personnel, with the corresponding safety measures.

We do real tests, on video! Watch them on our YouTube channel →

Do you have a problem nobody has been able to measure?

Talk directly to the engineer in charge, with no middlemen.

Talk to an engineer