The role of the antenna
A radiated emissions test does not measure power directly: it measures the field the equipment produces at a defined distance. The antenna receives that field and delivers a voltage to a measuring receiver. The antenna factor, given by its calibration and dependent on frequency, converts that voltage into field strength. The cable losses are also added to the result.
What is configured in the test
- Antenna type: to cover a wide range, broadband antennas are used, such as biconical, log-periodic or horn antennas, depending on the band.
- Measurement distance: defined by the standard applicable to the product, and fixed between the equipment and the antenna.
- Polarization: measurements are made in horizontal and vertical, because the equipment can emit in either.
- Antenna height: on sites with a ground plane, a height scan is performed to find the maximum of the sum of the direct and reflected signals.
- Equipment rotation: the equipment is rotated to find the direction of maximum emission.
An example workflow
The equipment is placed on a turntable and the antenna on a mast, and a frequency sweep is performed with a receiver. For each signal found, the value is maximized by varying height, polarization and angle, and then the final value, corrected for antenna factor and losses, is recorded.
How to interpret the result
The corrected level is compared with the limit in the product standard. A small margin indicates a risk that the measurement uncertainty will decide conformity, so the system uncertainty should be considered, which includes the antenna calibration, placement repeatability and the environment.
What to watch for
Reflections, ambient noise and expired calibration are the most common causes of doubtful results. Periodic site validation and antenna calibration are part of a reliable measurement.