Principle
According to Faraday’s law, a time-varying magnetic field passing through a loop induces a voltage in it. That voltage is proportional to the frequency, to the area of the loop and to the field component perpendicular to the plane of the loop. That is why a loop antenna is sensitive to the magnetic field, and more so at higher frequencies, while at very low frequencies it delivers very little signal.
What you get in practice
The antenna does not deliver the field directly: it delivers a voltage to a receiver or analyzer. To convert this into a magnetic field, the antenna factor is applied, which is determined in a calibration and supplied as a curve against frequency. The result is expressed as magnetic field strength or, through the plane-wave relationship, as its electric equivalent, although this equivalence is valid only in the far field.
Orientation and directionality
- Maximum response occurs when the axis of the loop, perpendicular to its plane, is parallel to the field.
- When the field is parallel to the plane of the loop, the induced voltage is zero.
- That is why the loop is rotated to find the maximum, or three orthogonal orientations are measured and combined.
Electric shielding
An unprotected loop also responds to the electric field, and this confuses the measurement. Many designs include a shield with a gap that interrupts the circulating current induced by the electric field but lets the magnetic flux through. The antenna therefore responds mainly to the magnetic field.
When it is used
- Measuring magnetic emissions close to sources, in the near field, where the magnetic field dominates for many low-impedance sources.
- Locating noise sources on boards, cables and equipment with small probes.
- Emissions tests at low frequencies, where magnetic components are measured.
What to watch for
The loop must be small relative to the wavelength; otherwise the current in the loop is no longer uniform and the antenna factor changes. Nearby metal objects and the position of the output cable alter the reading, so the same geometry should be kept during measurement and calibration.