What it is
A radio-frequency anechoic chamber is a shielded room, which keeps signals from entering or leaving, whose interior surfaces are covered with absorbers. It is called anechoic because it produces no echo: the energy reaching the walls is absorbed rather than reflected. The result is an environment equivalent to free space, where the received signal comes only from the direct path.
How the absorber works
The material usually takes the form of pyramids or wedges loaded with carbon or ferrite. The tapered shape provides a smooth transition between the air and the material and avoids the strong reflection a flat surface would produce. The energy is converted into heat inside the material. Its effectiveness depends on frequency, since the absorber must have a depth comparable to the wavelength to work well, so chambers for low frequencies are much larger.
What it is used for
- Radiation patterns: the signal is measured in each direction without echoes distorting the pattern.
- Gain and efficiency: the comparison method with a reference antenna requires a controlled field environment.
- Radiated emission and immunity tests: the shielding keeps external signals from contaminating the measurement.
- Cross-polarization: low values can only be measured if there are no reflections raising them.
What the engineer sees in practice
The quality of a chamber is verified with measurements of the quiet zone, the region where residual reflections are below an acceptable level. When the antenna is placed outside that zone, ripple appears in the pattern. The size of the antenna and the far-field distance determine whether it fits in the chamber or whether an alternative method is needed, such as near-field measurement or a compact range.
Limitations
No chamber is perfectly anechoic: residual reflections always remain, and at low frequencies the absorber loses effectiveness. The absorber also has power limits, because it can heat up if a lot of energy is transmitted.