How is the front-to-back ratio measured?

The front-to-back ratio is obtained from the radiation pattern: it is the difference, in decibels, between the level in the direction of maximum radiation and the level in the opposite direction, at 180 degrees.

Practical definition

The front-to-back ratio describes how much more an antenna radiates or receives toward the front than toward the back. It is calculated from a cut of the radiation pattern: the level at the maximum of the main lobe is taken and the level in the opposite direction is subtracted. A large ratio indicates that the antenna rejects signals arriving from behind well, which matters when interference or sources located in the opposite direction must be avoided.

How the measurement is made

  1. Place the antenna under test on a positioner, at a distance that satisfies the far-field condition relative to a probe antenna.
  2. Rotate the antenna in the plane of interest while recording the received level.
  3. Identify the maximum of the main lobe and the value at 180 degrees from it.
  4. Subtract the two levels in decibels.

What is seen in the pattern

The back lobe is usually much smaller than the main lobe, and its level can be sensitive to mechanical details such as the mast, the support, the cables or the antenna housing. That is why it is recorded with the mounting that will be used in the field. The minimum does not always fall at exactly 180 degrees, and the ratio depends on the definition adopted: some specifications use the value exactly opposite and others the largest lobe in a rear sector. It is advisable to state which one was used.

What limits accuracy

  • Reflections from the environment: at very low levels, an echo can be larger than the real signal from the rear.
  • Receiver dynamic range: it must allow the rear level to be distinguished from noise.
  • Cable leakage: currents on the shield can radiate and alter the rear pattern.
  • Alignment of the rotation axis: if the center of rotation does not coincide with the antenna, the pattern is distorted.

When it is not enough

The front-to-back ratio describes only two directions. To assess overall rearward rejection, the entire rear hemisphere must be examined in more than one plane.