How is gain determined by comparison?

The received signal is measured with an antenna of known gain and then with the antenna under test, in the same geometry. The difference in decibels between the two readings, added to the gain of the reference, gives the gain of the antenna under test.

The idea

Instead of measuring gain in absolute terms, a standard antenna of known gain is used. If everything else stays the same, the change in received power when one antenna is substituted for the other reflects the difference in gain between them. This is also known as the gain transfer method.

Calculation

In decibels, the gain of the antenna under test is the gain of the reference antenna plus the difference between the power received with the antenna under test and the power received with the reference. If the antenna under test receives less, the difference is negative and its gain is lower than that of the reference.

Procedure

  1. Mount the transmitting antenna and a receiving antenna, the reference, with the far-field distance and the polarization fixed.
  2. Precisely align the receiving antenna to the maximum and record the received power or the transmission between ports.
  3. Without changing the transmitting antenna, the power or the cables, replace the reference with the antenna under test, place its center in the same position, align to the maximum and record again.
  4. Correct for the mismatch of each antenna if their reflections differ appreciably.
  5. Calculate the difference and add it to the reference gain.

Conditions that must be met

  • The polarization of both antennas must be the same as that of the transmitting antenna.
  • The position of the phase center and the distance must be equivalent.
  • The environment must be the same for both readings and free of significant reflections.
  • The reference gain must be certified at the frequencies of interest. Its uncertainty carries over in full to the result.

Sources of uncertainty

In addition to the uncertainty of the reference, the factors include alignment error, the mismatch of each antenna, the stability of the source and receiver, reflections from the environment, and repeatability when changing antennas and connectors. Each one is estimated and combined to give a total value.

When it does not apply

If no antenna of known gain is available, the gain can be determined without a reference using the three-antenna method: the three pairwise combinations are measured and the system of equations is solved. The method is also poorly suited when the antennas differ greatly in polarization type, band or connector, because it is then difficult to keep conditions equal and the comparison loses validity.