How is an antenna tuned to a target frequency?

The reflection of the antenna is measured with a network analyzer, the resonant frequency is located and its electrical length is modified: shortening it raises the resonance and lengthening it lowers it. Corrections are made in small steps with the antenna installed.

Principle

The resonant frequency of an antenna depends on its electrical length: the greater the length, the lower the resonance. If the antenna resonates below the desired frequency, it is electrically too long; if it resonates above it, it is too short. Tuning consists of bringing that resonance to the target frequency and checking that, at that point, the impedance is close to that of the line.

How it is measured

  1. Calibrate the network analyzer at the end of the cable that connects to the antenna, over a band wider than the band of interest.
  2. Connect the antenna in its final mounting, or as close to it as possible.
  3. Observe the return loss: the resonance appears as a reflection minimum. On the Smith chart, the curve crosses the real axis where the reactance is zero.
  4. Note the frequency of the minimum and the reflection achieved there.

How it is corrected

  • Resonance too low. Shorten the radiating element in small sections and measure again after each cut. It is better to remove too little: material that has been trimmed cannot be put back.
  • Resonance too high. Lengthen the element, add a section or load it with a reactive element.
  • Right frequency but poor match. The resonance is fine, but the resistance does not match that of the line. This is adjusted by changing the feed point, the spacing to the ground plane or a matching network, not the length.

What shifts the resonance

The frequency measured with the antenna on the bench can change once it is installed. An enclosure, a shield, a hand, a metal mast, a ground plane of a different size or a plastic cover load the antenna and almost always lower the resonance. That is why final tuning must be done with the antenna in its real environment, or by applying a known correction.

Watch out for the cable

In antennas fed in an unbalanced way, such as a dipole connected directly to a coaxial cable, the outside of the cable can carry currents and behave as part of the antenna. The measurement then changes when the cable is moved. This is reduced with a balun or a choke, and you should verify that moving the cable does not change the trace.

How to know it is right

The minimum reflection must coincide with the desired frequency and its level must be sufficiently low; in addition, the impedance bandwidth must cover the band of interest with margin, because temperature, humidity and mounting shift the resonance slightly.