Why an anechoic chamber
The pattern describes the radiation of the antenna alone. In an ordinary space, walls, floor and objects reflect part of the signal, which adds to the direct signal and distorts the pattern. An anechoic chamber lines its surfaces with radio frequency absorbing material so that almost nothing is reflected, making the antenna appear to be in free space.
Setup
- Antenna under test. It is fixed to a positioner that rotates it in azimuth and, if required, on another axis. The support must be made of a low-reflection material and be aligned with the center of rotation, ideally the phase center of the antenna.
- Probe antenna. Fixed at the other end of the chamber, with known polarization. It illuminates the antenna under test.
- Instrument. A network analyzer measures the transmission between the two antennas, with the source at the probe port and the receiver at the antenna under test, or the other way around by reciprocity.
- Distance. It must be at least the far-field distance of the antenna under test. If that does not fit, compact range systems or near-field measurement with transformation are used.
Procedure
- Align the boresight of the antenna under test with the probe and set the angular zero.
- Calibrate the path with the cables and check the stability of the system.
- Rotate the antenna under test in fine steps or continuously, recording magnitude, and phase if required, by angle.
- Repeat for each cut plane and for each probe polarization: co-polar and cross-polar.
- Repeat at several frequencies in the band.
- If gain is required, compare against an antenna of known gain in the same configuration.
How to interpret the result
The curve is normalized to the maximum, and the beamwidth, side lobes, nulls and cross-polarization are read from it. If the maximum does not coincide with the expected axis, there is an alignment or mounting error. Rapid ripple in the pattern usually indicates residual reflections in the chamber or radiation from cables and supports.
What limits accuracy
- Quality of the quiet zone, the volume where residual reflections are tolerable.
- Radiation from cables that couple to the antenna; this is mitigated with ferrites and suitable routing.
- Misalignment of axes and polarization.
- System stability during the measurement, including cable drift during rotation.