Where noise comes from
Every conductor at a temperature above absolute zero generates noise through the thermal agitation of its charge carriers. This power is proportional to the temperature and to the bandwidth in which it is measured. At the usual reference temperature, the thermal noise power density is approximately minus 174 dBm per hertz. To this unavoidable noise the receiver adds its own, which is summed up in the noise figure: how many decibels the signal-to-noise ratio degrades between its input and its output.
The relationship that sums it up
The noise level referred to the receiver input, in dBm, is the sum of three terms: the thermal density (minus 174 dBm per hertz), ten times the logarithm of the measurement bandwidth in hertz, and the noise figure in decibels.
Direct consequences follow from this. Reducing the bandwidth by a factor of ten lowers the floor by ten decibels. A receiver with a lower noise figure has a lower floor. And an attenuator or a lossy cable ahead of the receiver raises the effective floor by the amount of its loss.
On a spectrum analyzer
The floor is seen as the displayed average noise level. It changes with the resolution bandwidth, the input attenuation and the use of the preamplifier. To measure weak signals, lower the resolution bandwidth, reduce the attenuation if it is safe to do so and turn on the preamplifier. The trace is smoothed with averaging or a video filter, but that does not lower the mean noise level.
On a network analyzer
The floor depends on the intermediate frequency bandwidth and the source power. It limits the ability to measure very large transmission losses, such as the rejection of a filter: beyond a certain value, the trace stops showing the response of the device and shows noise instead.
How to interpret a reading close to the floor
- A signal equal to the floor produces a reading about three decibels higher, because the noise power is added to it.
- A few decibels above the floor, the amplitude reading still has a significant error from the added noise.
- For good accuracy, a clear margin above the floor is needed, or the reading must be corrected for noise.
It is important not to confuse the noise floor with the dynamic range of the instrument, which also depends on its linearity and on the strong signal present at the same time.