What is the reference plane in a measurement?

It is the position along the signal path to which the results of a measurement are referred. Everything before that plane is corrected or excluded; everything after it is part of what is measured.

The basic idea

A network analyzer measures at its own ports, but what you normally want to know is the behavior of the device, not that of the cables, adapters and fixtures that connect it to the instrument. The reference plane is the location where the device is declared to begin. The measurement is corrected mathematically so that it appears to have been made exactly there.

How calibration establishes it

The reference plane is defined by the position at which the standards are connected during calibration. If the open, short and load are connected at the end of the test cable, that end is the plane. After calibration, the effect of the cable is removed and a device connected at that point is measured without it. If, instead, you calibrate at the instrument port and then add a cable, the cable becomes part of the measured device: its phase, loss and reflection appear in the results.

What happens when the plane is wrong

  • Transmission and reflection phase include the electrical length of any uncorrected section.
  • Insertion loss includes that of cables and adapters.
  • Additional reflections appear, producing ripple in the return loss.

This is especially delicate when the device connector does not match the test port connector and an adapter is needed, or when the device sits in a fixture, on a printed circuit board or on a wafer, where coaxial standards cannot be connected.

How to move the plane

Technique What it does Limitation
Port extension Subtracts a known electrical delay, and sometimes a known loss Corrects only a uniform, well-matched section
De-embedding Removes the effect of a fixture using its model or its measured S-parameters Requires good characterization of the fixture
Calibration with standards in the fixture Sets the plane directly where the device connects Requires standards to be fabricated in that medium

An example

If a filter with female connectors is measured using cables with female connectors, a male-to-male adapter between the cable and the filter enters the measurement if the calibration was done without it. The correct approach is to calibrate with the adapter installed, or to include its effect in the model.

A good practice is always to state where the reference plane is when reporting results, because without that information a phase or loss value cannot be interpreted.