How Is a Limiter or Input Protector Tested?

It is verified in two regimes: with a small signal, to confirm that it has almost no effect on the useful signal, and at high power, to measure the level at which it starts limiting, how much power it lets through and how long it takes to recover.

What a limiter must do

A limiter protects the input of a receiver. With weak signals it must be almost transparent, and when a strong signal arrives it must clip it to a level the receiver can tolerate. Common limiters use diodes, so their behavior depends on power level and is not linear. That requires testing in more than one regime.

Small-signal test

With a network analyzer at low level, insertion loss and return loss are measured in the band of use. This part confirms that the limiter does not degrade the receiver’s sensitivity or matching. High insertion loss here directly penalizes the receiver’s noise figure, because the limiter is usually placed ahead of the low-noise amplifier.

High-power test

A single-tone signal of increasing level is applied and the output power is recorded. The resulting curve starts out linear, bends at the limiting threshold and then flattens. The parameters of interest are:

  • Limiting threshold. The input level from which the output stops following the input.
  • Flat leakage. The power that keeps passing once the limiter is fully active, which is what the receiver must withstand.
  • Spike leakage. With pulses, the initial part of the pulse passes with greater amplitude before the limiter reacts. It is observed with an oscilloscope and a detector, or with a peak power meter.
  • Recovery time. The time the limiter takes to return to its normal loss after a strong event, during which the receiver is desensitized.
  • Harmonics. Clipping generates harmonics, which are measured with a spectrum analyzer.

Setup and precautions

The high-power source is connected through suitable attenuators and couplers so that the measuring equipment receives a safe level. Loads and cables must withstand the applied power. Raise the level carefully and monitor the device temperature, because a limiter held at its limit for a long time can be damaged. Pulsed and continuous-wave tests give different results, so the actual conditions of use must be reproduced.

Work with high radio-frequency power involves a risk of burns and exposure. It must be carried out by qualified personnel following the applicable regulations.