How Is a Power Divider Evaluated?

A network analyzer is used to measure insertion loss, amplitude and phase balance between outputs, isolation between them and the match at each port, always terminating the ports not being measured with loads.

What is expected of a divider

An ideal two-output divider splits the input power equally, so each output receives half, about three decibels down. That split is not loss: it is the division itself. A real divider also adds its own losses, imbalance between outputs, coupling between them and reflections. The evaluation quantifies those deviations.

Parameters measured

Parameter What it indicates How it is obtained
Insertion loss Additional loss relative to the ideal split Transmission from the input to each output minus the theoretical division
Amplitude balance Level difference between outputs Difference between the two transmissions
Phase balance Phase difference between outputs Difference between the phases of the two transmissions
Isolation How much signal passes from one output to the other Transmission between the two outputs
Return loss Match quality at each port Reflection at the input and at each output

Procedure

  1. Calibrate the analyzer at the plane of the divider’s connectors, with the connector type it has.
  2. Connect the input and one output, and terminate the other with a good-quality load; an unterminated output reflects and alters all the readings.
  3. Measure and record. Repeat with the other output.
  4. Measure the isolation between the outputs, now with the input terminated.
  5. Measure the match at each port, likewise with the others terminated.

If the analyzer has only two ports, repeat with different combinations and loads. With an analyzer of three or more ports, all combinations can be measured in one sweep.

How to interpret the result

A resistive divider and a Wilkinson divider differ in what matters. The resistive type is broadband and simple, but its loss is higher and its isolation is modest. The Wilkinson type has an internal resistor between the outputs that improves isolation and keeps all ports matched when the loads are correct, but its good behavior is limited to one band. If the measured isolation is low at a given frequency, or the phase between outputs deviates, the effect will appear in any system that combines or compares those outputs, for example in antenna arrays.

The maximum allowable power, especially that of the internal resistor, must also be verified before using the divider with strong signals.