What torque does in a coaxial connector
In a threaded connector, the nut presses the outer conductors together and ensures that the center conductors sit at the correct depth. That pressure must be enough for the contact to be continuous around the entire circumference, but not so much that it deforms the parts. That is why manufacturers specify a torque for each connector type, applied with a torque wrench of the appropriate size.
What happens with too little torque
- The contact between outer conductors is not uniform, and a small, variable resistance appears, with signal leakage through gaps.
- The joint may have axial play, which changes the position of the center contact and the reflection.
- Each time it is connected the result is different, so the calibration and the measurement do not correspond.
What happens with excessive torque
- The mating faces and the center contact deform, and the damage is permanent.
- The depth of the contacts relative to the reference plane changes, so the connector ends up out of tolerance and reflects more.
- Wear on the threads and on the instrument connectors, which are costly, accelerates.
What you see in the trace
The most characteristic sign is lack of repeatability: when the same device is disconnected and reconnected, the return loss changes more than it should with a healthy connector. With too little torque, the trace may also vary when the cable is touched. In a calibrated sweep, a different torque between calibration and measurement shows up as a systematic error that no correction removes, because the calibration was made under a different connection condition.
What to check first
- That a torque wrench for the connector type is used, in good condition and periodically verified.
- That only the nut is turned, holding the body of the other connector, so as not to twist the center contact.
- That tightening is smooth until the wrench releases, without jerks.
- That the torque is the same in calibration and in measurement.
- That the connectors are clean and undamaged; a damaged connector cannot be corrected with torque.
The correct torque value depends on the connector type and must be taken from the manufacturer’s specification.