Why it happens
The first stage of a spectrum analyzer, normally a mixer, is linear only up to a certain input level. If the signal reaching that stage is higher, the mixer compresses: its gain falls and it distorts the signal, just as a saturated amplifier does. The distortion generates harmonics and intermodulation products that the analyzer displays as if they were signals present in the circuit under test.
What matters is that the relevant level is the total power reaching the mixer, not only that of the signal you want to measure. A strong signal outside the observed band, or several signals, can saturate the input even if the displayed region looks quiet. In analyzers with digital conversion, the converter can also saturate, and the instrument usually warns of this with an overload indicator.
How to recognize it
- Unexpected harmonics or spurious products appear.
- An overload or input-level-exceeded indicator appears on the display.
- The level of the main signal seems lower than expected.
The clearest test is to increase the input attenuation and see how the spurs respond. If they are real, they fall by the same amount as the main signal. If the analyzer itself generates them, they fall more than the signal, because distortion products grow faster than the signal that produces them. A spur that changes disproportionately with attenuation is an instrument artifact.
How to avoid it
- Increase the input attenuation until the spurs disappear or stabilize, accepting a higher effective noise floor.
- Set the reference level so that the maximum signal sits near the top of the display; on many instruments this also sets the automatic attenuation.
- Install an external attenuator, a filter that rejects strong signals outside the band of interest, or a coupler when the source has high power.
- Check the instrument’s maximum input power limit; exceeding it can damage the mixer and not merely saturate it. Also verify that there is no DC voltage at the input.
When measuring high-power transmitters, a coupler or attenuator with the correct power rating must be used, and the work must be done by qualified personnel following the applicable safety regulations.