What a noisy trace means
The analyzer always displays its own thermal noise and that of its electronics. When the measured signal is well above it, the noise goes unnoticed. As the signal gets closer, the trace becomes irregular and the reading fluctuates. The useful question is how much noise enters the detector and how much of it is averaged.
The settings with the most influence
- Resolution bandwidth (RBW). The noise power the instrument sees is proportional to this width. Reducing it by a factor of ten lowers the displayed noise level by ten decibels. In exchange, the sweep becomes slower.
- Video bandwidth (VBW). This is a post-detection filter. Reducing it smooths the trace but does not change the average noise level. If the trace looks irregular because of rapid fluctuations, this is the setting to adjust.
- Averaging. Averaging several sweeps reduces trace variation. Power averaging is preferable when measuring noise.
- Input attenuation. High attenuation protects the mixer but raises the effective noise floor. If the signal is small and safe, it can be reduced.
- Preamplifier. If the instrument has one, it lowers the effective floor for small signals, at the cost of less margin against strong signals.
- Detector. The detector type changes how noise appears: a peak detector shows it higher than an average detector.
Causes unrelated to the settings
A defective cable, a loose connector or a connection with a poor ground introduce noise and variations. The environment also has an influence: nearby emissions or an input antenna pick up signals that are interpreted as noise. To tell them apart, remove the signal and place a load on the input; if the trace cleans up, the origin is external.
An example adjustment sequence
- Connect a load and observe the real floor with the current settings.
- Lower the RBW until the signal of interest stands out clearly, accepting a slower sweep.
- Reduce the VBW and, if necessary, average to stabilize the reading.
- Review the attenuation and the preamplifier without putting the input at risk.
Keep in mind that if the signal is within a few decibels of the floor, the reading includes noise, so its amplitude must be interpreted with care.