What a load does and why it matters
A transmission line behaves as expected only when it ends in its own characteristic impedance. If the end is different, part of the wave returns and overlaps the incident wave. The coaxial termination load solves this: inside the connector there is a resistive element, usually a resistive film, sized to match that impedance and convert the incident power into heat.
In terms of parameters, an ideal load has a reflection coefficient of zero, that is, infinite return loss. No real load achieves this: a small residual reflection always remains, and that figure is what distinguishes an ordinary load from a precision one.
Where it is used
- Calibration standard. On a network analyzer, the load is one of the standards, along with the open and the short. The known reflection coefficients of the three standards make it possible to calculate and correct the instrument’s systematic errors.
- Termination of unused ports. On couplers, dividers, circulators or switches, free ports are terminated so that they do not reflect and alter the measurement.
- Power dissipation. Some loads are designed to absorb considerable power and include a heat sink.
What you see in practice
If a load is connected to a calibrated port and the reflection is measured, the return loss trace is good at low frequency and deteriorates as frequency rises. The cause is that at high frequencies the parasitic inductance and capacitance of the resistive element and its transition to the connector are no longer negligible. That is why the specification of a load always depends on frequency.
In a calibration, the load defines the analyzer’s directivity: what the instrument measures as reflection when it should be zero. A load of poorer quality than the calibration model assumes directly limits the accuracy of every later reflection coefficient measurement.
What distinguishes one load from another
| Characteristic | Effect |
|---|---|
| Return loss | How closely it approaches zero reflection in each band |
| Power rating | Thermal limit before the resistive element is damaged |
| Connector type and gender | Must match the port; adapters add reflections |
| Sliding load | A variant that is moved to separate out its residual reflection and achieve a better reference at high frequency |
Precautions
A load must not be confused with an attenuator: an attenuator has two ports and lets part of the signal through. Its rated power must not be exceeded either, because heat permanently alters the resistance and the load stops being a reliable standard. To use it as a calibration standard, use the model or coefficients that correspond to the specific calibration kit.