What Is a Power Meter with a Sensor Used For?

It measures, accurately and with traceability, the absolute power reaching a sensor, regardless of frequency selectivity; the sensor converts RF power into an electrical signal and the meter processes and displays it.

Function

A power meter measures the RF or microwave power absorbed by a sensor connected to its input. It is the reference instrument for determining the absolute level of a signal with good accuracy, and for calibrating signal generators, amplifiers, attenuators and other laboratory equipment.

How the tasks are divided

  • The sensor is the element that absorbs the power and converts it into an electrical signal. It can be thermal or diode-based, and each type covers a given frequency and power range. It contains its own calibration data.
  • The meter powers the sensor, amplifies and processes the signal, applies frequency correction, zeroing and calculations, and presents the result in watts or dBm.

What it offers compared with a spectrum analyzer

The meter has better absolute level accuracy and better repeatability, and its calibration is performed against its own reference. On the other hand, it is not selective: it sums all the power reaching the sensor within its range, including harmonics and spurious signals. If you need to see how much power each component has, the analyzer is the tool. If you need the most accurate absolute figure for a single signal, use the meter.

Common uses

  • Verifying the output power of a transmitter or generator.
  • Calibrating the level of a signal generator.
  • Measuring amplifier gain by comparing input and output power.
  • Characterizing pulses, with peak power meters and sensors.

What you should know

  • The sensor has a maximum power that must not be exceeded; attenuators are used for high power.
  • Mismatch between the source and the sensor introduces an uncertainty that must be taken into account.
  • Before measuring, zero the meter with no signal applied and enter the frequency so that the correct calibration factor is applied.

How to interpret the result

The value is the power absorbed by the sensor, with an associated uncertainty made up of sensor calibration, mismatch, noise and linearity.