How is the isolation between ports of a divider measured?

By measuring the transmission between the two output ports with the input terminated in a matched load and the remaining ports loaded; the level of that transmission, in decibels, is the isolation.

What isolation means

In an ideal three-port power divider, a signal applied to one output must not appear at the other. Isolation measures how much this transfer between outputs is attenuated. A simple resistive divider does not offer good isolation, whereas types with an isolation resistor, such as the Wilkinson divider, improve it in the band for which they are designed, provided the ports are properly terminated.

How it is measured

  1. Calibrate the network analyzer in transmission with the reference planes at the output connectors.
  2. Connect one output port to each analyzer port.
  3. Terminate the divider’s input port with a matched load of the same impedance as the system. This is essential: an open or poorly terminated input changes the result.
  4. Measure the transmission between the two outputs, for example S32 if the outputs are ports 2 and 3.

The match of each port is also measured to have the full context.

Limitations

  • Dynamic range. High isolation means a small signal, close to the noise of the analyzer’s receiver. The intermediate-frequency bandwidth is reduced and averaging is applied to improve the measurement.
  • Leakage between cables and connectors. Low-quality cables leak signal and limit the measurement.
  • Load quality. Any reflection from the input termination is reflected back and alters the result.

How to interpret the result

Isolation varies with frequency and is usually better near the center of the design band. A poor value can mean a damaged isolation resistor, mismatch at the ports or a defective termination. A divider used as a combiner needs good isolation so that the sources do not interact with each other.