What it is
The resolution bandwidth (RBW) is the width of the intermediate-frequency filter through which the signal passes before detection. It is the filter with which the analyzer observes the spectrum, and its shape appears on the screen in every signal.
What changes when you modify it
- Resolution. Two signals closer together than the RBW merge into a single peak. To separate them, an RBW smaller than their spacing must be used.
- Noise floor. The noise passing through the filter is proportional to its bandwidth. Reducing the RBW by a factor of ten lowers the displayed noise by ten decibels. This helps in seeing weak signals. A sinusoidal signal does not change level, because all its power fits within the filter.
- Sweep time. A narrow filter needs time to respond. In a swept analyzer, the minimum time grows approximately with the span divided by the square of the RBW. Reducing the RBW by a factor of ten can lengthen the sweep about a hundred times.
- Filter shape. Its selectivity, the ratio of the width at a low level to the width at three decibels, determines whether a weak signal next to a strong one is visible or masked.
How to decide
- Set the span needed to see what is of interest.
- Choose an RBW smaller than the spacing between the signals you want to distinguish.
- If the noise floor hides signals, reduce the RBW until they appear, accepting a slower sweep.
- If the sweep is too slow, increase the RBW or reduce the span.
A practical case
When looking for modulation sidebands around a carrier, the RBW must be well below the modulation frequency. When measuring the power of a wide channel, a large RBW reduces the sweep time without losing relevant information.
What you should know
If the RBW is too large for the span, or the sweep too fast, a warning appears that the measurement is uncalibrated, and the displayed level is incorrect. When measuring noise, the value must be normalized to the bandwidth and the type of detector taken into account.