Calibrating is not the same as verifying
A calibration builds an error model from standards. Verification checks, with an independent element, that the model works. If verification is done with the same standards used for calibration, the result always comes out well by construction, without providing any information.
Which standards work
- Verification kit. Devices with certified reference data, such as attenuators, lines and known mismatches.
- Air line. Its impedance depends only on dimensions, so it has a calculable response. It makes it possible to check directivity and source match.
- Characterized attenuators and adapters. Useful for checking transmission.
- A stable laboratory device. A filter or cable of your own, measured repeatedly, serves to monitor drift between calibrations even if it has no absolute data.
How it is done
- Perform the calibration with the normal procedure.
- Connect the verification standard without recalibrating.
- Measure the relevant S-parameters over the frequency range.
- Compare with the reference data, plotting the difference or the uncertainty limits.
How to interpret the result
The differences must fall within the combined uncertainty of the standard and the system. A pattern of differences that oscillates with frequency points to poor directivity or match, often due to dirty, worn or poorly tightened connectors. A constant difference in transmission may indicate a damaged cable or a poorly defined thru standard. If the system fails, the most likely cause is a connection problem, not the instrument, and it is advisable to clean, inspect the connectors and repeat.
When to repeat it
At the start of an important session, after moving the cables, after a noticeable temperature change or when a result seems suspicious. Recording verification results over time shows the drift of the system.