How is a network analyzer calibrated with the SOLT method?

Four known standards (short, open, load and direct connection) are measured so that the analyzer can calculate the systematic errors of the instrument and cables and remove them mathematically from every subsequent measurement.

What a calibration corrects

A network analyzer measures with repeatable systematic errors: losses and mismatches in cables and connectors, leakage between channels and signal leakage. A vector calibration characterizes these errors by measuring standards of known behavior and subtracts them from every measurement. As a result, the reference plane moves from the instrument connectors to the end of the cables, where the device is connected.

The SOLT method

The name comes from the standards: Short, Open, Load (a matched load) and Thru (a direct connection). It is suitable for coaxial connectors for which good-quality standards exist.

  1. The calibration kit to be used, with the characteristics of its standards, is defined in the instrument.
  2. The short, open and load are connected in turn at each port, and the analyzer measures the reflection of each one. They determine three errors per port: directivity, source match and reflection tracking.
  3. The thru is connected between the ports and the transmissions are measured. This determines the transmission tracking and load match errors.
  4. Isolation measurements are completed if the case requires it.

Together these yield a twelve-term error model for two ports.

What determines the quality

  • The accuracy of the standard definitions: a poorly described kit introduces direct errors.
  • The quality of the load, which sets the residual directivity and limits the measurement of small reflections.
  • The repeatability of the connections: tightening to the correct torque, with clean, unworn connectors.
  • Cables at rest: moving them after calibration changes their response.

When it does not work well

If the device connectors cannot be mated directly (an adapter must be used), the thru is not perfect and another method is advisable. In non-coaxial media, such as microstrip or on-wafer, where precise standards are lacking, methods such as TRL are usually preferred.

What to check afterward

Measure a known standard and confirm that the result agrees, as verified in another guide on calibration verification.