The principle
Two impedances in series divide the voltage according to their ratio. A high-voltage arm withstands nearly all of the voltage and a low-voltage arm delivers the fraction that is measured. The ratio of the divider is the quotient of the total voltage and the measured voltage. If the impedances maintain their proportion at all frequencies of interest, the reduced signal faithfully reproduces the shape of the original voltage.
Types and what they are for
- Resistive: very well suited to DC and low-frequency voltage. At high voltage it dissipates power and can heat up, and stray capacitances limit its high-frequency response.
- Capacitive: the most widely used for AC voltage and impulses, because it consumes no active power and has a good frequency response. It is sensitive to stray capacitance to ground and to nearby objects.
- Mixed resistive-capacitive: combines both to maintain the division ratio over a wide frequency range, typical of impulse measurements.
What determines measurement quality
- Frequency response: for impulse waves, which have fast components, the divider must respond in very short times.
- Stray capacitances: those from the high-voltage arm to ground and to nearby structures distort the voltage distribution. For this reason, attention is paid to the distance from other objects and shielding rings are used.
- Connecting cable: it must have a matched termination to avoid reflections.
- Temperature and voltage coefficient: the impedances must not change appreciably during the test.
How to interpret the result
The voltage read is multiplied by the divider ratio, which is determined by calibration against a reference system. A divider that has not been verified can introduce a silent error into the entire test. Calibration must be repeated periodically.
These activities involve hazardous voltages: they must be carried out by qualified, authorized personnel, following the applicable regulations and the facility’s lockout and grounding procedures.