Why data is aggregated
An analyzer measures continuously, but storing every cycle for weeks is impractical and hard to interpret. Aggregation reduces the data: it calculates quantities over very short windows of an integer number of cycles and combines them into longer intervals, which are what is stored, plotted and compared with criteria.
How it is done
- Basic window. The instrument calculates the RMS value, or the harmonics, over an integer number of cycles of the line frequency. Using complete cycles avoids truncation errors.
- Combination. The basic windows of an interval are combined. For quantities such as RMS voltage, the root mean square is used rather than the arithmetic mean, because signal energy is proportional to the square.
- Recording interval. The long interval, defined by the applicable criterion (for example, ten minutes for voltage, or the demand interval for power), is the one evaluated.
- Additional values. Each interval can store the minimum and maximum of the basic windows and, depending on the instrument, percentiles.
What is lost by averaging
A voltage sag of a few cycles, a transient or a brief current peak barely moves the average of a long interval. That is why events are detected in parallel, with thresholds on the half-cycle RMS value or on the waveform, and are recorded with their own date and duration, without passing through the average. The minimum and maximum values of the interval are the sign that something happened within the average.
What an engineer sees in practice
Two records from the same point with different aggregation intervals will show different maximum values: the short interval shows higher peaks; the long one, a smoother value. That is why a result is compared with a limit only if both use the same aggregation. The instrument may also flag or discard intervals in which an event occurred, so as not to contaminate the statistics.
How to interpret the result
To evaluate compliance, take the aggregated values and calculate the required percentile over the period. To diagnose faults, turn to the minimums, maximums and waveforms. Always check that there are no data gaps and that the time is synchronized.