Measurements and testing of dielectric mats and carpets in Monterrey

We test dielectric mats and flooring at our laboratory in Monterrey. For anyone working with static-sensitive electronics who needs to know whether their mat still works.

What dielectric mats are for

Static electricity on your hands or in the work area can destroy metal-oxide semiconductor transistors and other electronic parts in an instant. Specially treated plastic mats, called electrostatic discharge or ESD mats, are placed on the bench and safely remove static charges while you work. They drain static electricity without short-circuiting components.

Contamination and wear can increase a mat’s resistance and compromise its performance. That is why it must be tested periodically to find out whether it needs to be cleaned or replaced.

An antistatic mat uses a conductive plastic material to drain static charges away from the work surface. The electrical resistance of its surface is usually in a range of 1 megohm (one million ohms) to 10 gigohms (one billion ohms). A typical electronics store multimeter does not measure resistances this high; specialized equipment stores sell meters capable of testing antistatic mats. Over time, accumulated dirt interferes with the mat’s resistance.

How a mat is tested

Grounding

Connect the mat’s ground clip to electrical ground. You can use a grounded metal conduit or a cold water pipe.

Measuring with the mat meter

Place the mat on a clean, flat, non-conductive surface. Turn on the test meter. Clamp one probe to the mat’s metal snap or grounding point. Touch the other probe to the point on the surface farthest from that point.

Read the display. Some meters have a resistance indicator and others have “good”, “high” and “low” resistance lights. A mat in good condition reads between 1 megohm and 10 gigohms, or lights the “good” light; it is also considered good if it reads a resistance between 1 million and 10 billion ohms.

Edge-to-edge measurement

Move the 2-meter probes to opposite ends of the mat, touching the surface. The meter should read a steady resistance of more than 1 million ohms. Move the probes to another point on the edge, always on opposite sides, keeping the full width of the mat between them. The meter should still show a resistance between 1 million and 10 billion ohms, or light its “good” light.

If the readings are not consistent, clean the mat with a mild cleaning agent and test it again. Periodic testing with the right meter will tell you when to clean or replace the mat.