Description
The interfacial tensiometer is applicable to the GB/T6541 standard. The force between molecules forms the interfacial tension or surface tension of the liquid. The tension value reflects the physical and chemical properties of the liquid and its composition, and it is one of the important indicators for inspecting product quality in the related industries.
The PS-ZL203 is based on the ring method (platinum ring). It measures the surface tension of various liquids (liquid-gas phase interface) and the interfacial tension of liquids (liquid-liquid phase interface).
Functions and features
- Original fast-response electromagnetic force balance sensor, which improves lateral accuracy and linearity.
- Instrument calibration is single-point. This resolves the shortcoming of the previous sensor generation, which needed multipoint calibration, and eliminates the zero and full-scale potentiometers.
- The equivalent tension value and current weight are displayed in real time.
- Integrated temperature sensing circuit, with automatic temperature compensation of the measurement result.
- 240*128 dot matrix LCD screen, with no identification key, with screen saver function.
- Time-stamped history log, with up to 255 records stored.
- Built-in high-speed thermal printer, with offline printing function.
Main technical parameters
| Measurement range | 0-200mN |
| Accuracy | 0.1% of reading ±0.1 mN/m |
| Sensitivity | 0.1 mN/m |
| Resolution | 0.1 mN/m |
| Supply voltage | 220 V AC ±10% |
| Supply frequency | 50Hz ±2% |
| Power | <20w |
| Applicable temperature | 10~40 ℃ |
| Applicable humidity | <85% RH |
| Width x height x depth | 200×330×300 (mm) |
| Net weight | ~ 5 kg |