Deep vacuum and controlled temperature
We combine deep vacuum and temperature control using high-strength acrylic chambers, high-vacuum pumps and Sun Electronics Systems climate control equipment. This recreates extreme conditions that reveal hidden failures, critical behaviors and structural limits in materials, electronic components, adhesives, packaging and mechanical systems.
How we perform the test
Progressive vacuum
We apply progressive vacuum inside the acrylic chamber. Atmospheric pressure is removed and the sample experiences differential stress in a controlled environment. This is useful for identifying internal collapses, layer detachment, manufacturing defects, hermeticity failures and changes in mechanical properties. The transparent acrylic allows real-time observation of deformations, bubble formation, invisible leaks and any alteration of the material.
Temperature
At the desired vacuum level, we use a Sun Electronics Systems environmental chamber that changes the temperature from extreme cold to high heat. When thermal shock must be simulated, liquid nitrogen is injected. These combined conditions make it possible to study thermal expansion, weakening points, adhesive strength, coating stability, the reaction of composite materials and the performance of electronic parts under severe stress.
Data logging
With high-precision digital instrumentation we record internal pressure, vacuum level and thermal variations in real time. From this we generate behavior graphs, pressure-temperature curves and detailed analyses that serve for audits, supplier validation, quality control, pre-certification testing and new product development.
Reports
The reports include photographic evidence and/or videos throughout the process, evidence of deformation, measurement records, behavior curves and technical conclusions. We work with transparency, traceability and complete visual documentation of each test.
Industries
The tests are suitable for the electronics, automotive, aerospace, polymer, export packaging and medical device industries, and for companies that need to validate the integrity of their products under extreme conditions.
Metrological temperature control
In our environmental analyses and laboratory tests where temperature is a critical parameter, we verify our thermal systems in-house. Each test is backed by temperature validation in our dry-well calibrator:
- Operating range of -30 °C to 150 °C.
- Resolution of 0.01 °C.
- Thermal stability of ±0.02 °C.
- Indication accuracy of ±0.1 °C.
- Block uniformity (horizontal ≤ ±0.05 °C) and controlled vertical gradient.
In this way we confirm the stability, uniformity and accuracy of the test conditions, with traceable and technically supported results. The expanded uncertainty of the system is determined according to the operating point, the sensor under test and the process conditions.