Oscilloscope: AC, DC and GND Coupling

Explanation of the oscilloscope AC-GND-DC selector and of what coupling is, for those starting to use an oscilloscope.

Oscilloscope: AC, DC and GND coupling

What coupling is

Coupling is the name given to a device or method whose purpose is to transfer energy. Some couplings can be deactivated during operation, which interrupts the energy transfer.

Oscilloscope AC-GND-DC selector (CRO)

The AC-GND-DC selector is very important because it lets you view the signal being measured as needed. There is one per channel and it has three positions: AC, GND and DC.

Most signals (waveforms) to be measured have a direct current (DC) component and an alternating current (AC) component. Sometimes you want to see only the AC component and sometimes you do not.

AC position

Shows only the alternating current component of the signal and removes the DC component, if there is one. To achieve this, a large-value capacitor in series with the input of each channel blocks the DC component. Remember that a capacitor does not pass direct current.

GND position

Disconnects the oscilloscope inputs. The screen shows a horizontal line that sets the zero (0) volt level.

It is useful for setting the zero-volt level wherever convenient on the screen, using the vertical trace adjustment knob. For example, instead of having the trace in the center of the screen, as is customary, you can raise or lower it as desired.

DC position

The signal is displayed exactly as it is, as a combination of AC and DC.

Keep in mind that the DC component of the signal, which is removed in AC measurement, can have a large value and push the signal off the screen. To solve this, adjust the vertical measurement scale appropriately.