Verifying the ground connection
In electrical safety testing, verifying a correct ground connection is indispensable for protecting the user. The ED&D GC-100 evaluates this condition with high-current tests and detects high resistances or poor contacts that could compromise the safety of the equipment.
Main functionality
This tester performs continuity tests by applying 25 A or 30 A, in line with applicable IEC standards. It includes automatic functions that simplify operation:
- Selectable test times (1 s or 60 s)
- Automatic evaluation with fail indication
- Lead compensation through a zero function
Key specifications
| Parameter | Specification |
|---|---|
| Test current | 25 A / 30 A |
| Test time | 1 s / 60 s |
| Fail threshold | 0.1 Ω (0.2 Ω optional) |
| Accuracy | ±2% |
| Power supply | 100–120 / 220–240 VAC |
| Weight | 15 lb |
Applications
- Validation of electrical and electronic equipment
- Testing on production lines
- Regulatory compliance testing
- Electrical safety laboratories
Benefits of renting
Renting the GC-100 gives you high-current testing without a permanent investment, and makes audits, certifications or specific projects easier.
- Fast deployment
- Reliable results
- Fewer operating errors
- Optimized costs