778D dual directional coupler rental, 100 MHz-2 GHz, type N connectors

20 dB dual directional coupler for swept measurements from 100 MHz to 2 GHz, for rent in Monterrey or Mexico.

HP 778D dual directional coupler, 100 MHz-2 GHz

Dual directional coupler for swept measurements

We rent the 778D in Monterrey or Mexico. It is a dual directional coupler designed for swept measurements in the 100 MHz to 2 GHz range. Its high directivity and its mean coupling accuracy of ±0.5 dB make it ideal for reflectometer applications. The auxiliary arms track very closely, which makes it particularly useful for reflectometers driven by sweep oscillators.

Broadband 20 dB dual directional coupler

This 20 dB dual directional coupler has a frequency range of 4 octaves, from 100 MHz to 2 GHz. Its multi-octave capability makes setting up measurements easier and reduces calibration time, since several couplers are normally required in this band.

The coupling factor increases by about 6 dB per octave below 100 MHz. For this reason, usefulness below this frequency is limited only by the power of the signal source or by the sensitivity of the RF detector connected to the coupler.

Impedance measurement

The 778D is also well suited to impedance measurements when used with the HP 8405 A vector voltmeter. The technique is described in Application Note 77-3, complex impedance measurement.

RF performance

  • Nominal coupling and variation: 20 ±1.0 dB
  • Directivity: >36 dB at 1 GHz
  • Maximum SWR: 1.1
  • Insertion loss: <0.60 dB

Power ratings

Average, peak rated power: 50 W, 500 W